Product Features | |||||
Medical & ITE safety meet | |||||
2 MOPP input to output isolation | |||||
Efficiency efficiency level V | | | |||
Touch current ≤ 100μA | |||||
≤0.3W standby power | |||||
Up to 5000m operating altitude |
Input |
Input Voltage Range:90-264VAC
Frequency Range:47-63Hz
Input Current:0.5A max at 100VAC
Inrush Current:60A max at 240VAC cold start
Touch Leakage Current(max):≤ 100μA at 264VAC
Environmental
Operating Temperature:-10℃ to 70℃(50℃ to 70℃ refer to derating curve)
Storage Temperature:- 20℃ to 70℃
Operating Humidity:10 to 90% RH,non-condensing
Storage Humidity::5 to 90% RH
Operating Altitude:5,000 M
General
Dimensions:65(L)×35(W)×24(H)mm
Weight:54g
MTBF:>100000hrs MIL-HKBK-217 at 25℃
Safety Approvals | |||||||||
Safety Agency/Mark | Medical(Meet) | I.T.E.(Meet) | |||||||
CB | IEC60601-1 | IEC62368-1 | |||||||
UL | | ANSI/AAMI ES60601-1 | | UL60950-1 UL62368-1 | |||||
| | CAN/CSA C22.0 NO.60601-1 | | CAN/CSA C22.0 NO.60950-1 | |||||
TuV Rheinland/Mark | | EN60601-1 | | - | |||||
TuV Rheinland/GS | | - | | EN60950-1 EN62368-1 |
Model Number | Voltage (V) | Current (A) | Rated Power (W) | Ripple & Noise (max) mVpk-pk | Voltage Tolerance | Line & Load Regulation | Efficiency | Start Up delay |
UE20-XXXYYYSPA-OP | 9.0 | 1.67 | 15.03W | 150 | ±5% |
Line: ±1% Load: ±5%
| 79.16% | ≤3s |
12.0 | 1.67 | 20.04W | 200 | ±5% | 80.97% | ≤3s | ||
15.0 | 1.34 | 20.01W | 200 | ±5% | 80.98% | ≤3s |